SW-9OOO is an Optical Biometer capable of highly accurate measurements along the visual axis and IOL calculation.
Technology overview: 820nm Superluminescent Laser Diode combined with Optical low-coherence Reflectometry(OLCR) allows capture of 8 measurements with one scan, and it has advantages as Fast Measurement, High Accuracy, Non-Contact and Comfortable etc.
| Measurement range | |
| Axial length | 12-34 mm |
| Çentral corneal thickness | |
| Corneal radii | 4.8-11.1mm |
| Axis angle | 0-180" |
| Anterior chamber depth | 1.5-6.0mm |
| Lens thickness | 0.5-7.0mm |
| White-to-white | 6.5-16.6mm |
| Pupil diameter | 1.9-13.5mm |
| Resolution | |
| Axial length | 0.01mm |
| Central corneal thickness | 1um |
| Corneal radii | 0.01mm |
| Axis angle | 1" |
| Anterior chamber depth | 0.01mm |
| Lens thickness | 0.01mm |
| White-to-white | 0.01mm |
| Pupil diameter | 0.01mm |
| SD of repeatability | |
| Axial length | t25um |
| Central corneal thickness | +2um |
| Corneal radii | +10um |
| Axis angle | +9® |
| Anterior chamber depth | t20um |
| Lens thickness | t50um |
| White-to-white | t0.3mm |
| Pupil diameter | +0.3mm |
| IOL calculation formulas | BinkHorst-ll, Holladay, Hoffer-Q,Haigis, SRK-T, SRK-II |
|
Calculation for eyes following refractive surgery: |
Shammas-PL, Masket, Modified Masket |
| Interfaces | USB2.0 |
| Voltage/Frequency | AC 220V/50Hz |
| Power consumption | 50VA |
| Laser class | 1 |