| Amplifier | |
| Input Impedience | ≥ 1000 MΩ |
| CMRR | ≥ 115dB |
| Input Noise | ≤ 4uVpp |
| Filter |
Full real time digital filter, band-pass, band-stop, high-pass, low-pass, comb filter apply to every channel independently |
| A/D Resolution | 24 bit |
| Port | Optical isolated USB 2.0 |
| Power Supply |
Li-ion battery, switch to USB power automatically when low battery |
| Flash Stimulator | |
| Light Source Color | White, Red, Green, Blue, Yellow and any of their combination |
| Flash Luminance | Can produce 100cd·s/m2 broad-band white flash in < 4ms |
| Flash Luminance Range | 0-1000 cd·s/m2 |
| Stimuli Mode | Single, Double, String, On/Off, On duration can be adjusted |
| Background Illumination | 0-3000 cd/m2 |
| Stimuli Frequency | 0.01-200Hz |
| Fixation & EOG |
11 Fixation/EOG indicator, 64 adjustable intensity; -25 - +25 EOG saccade range |
| Monitor | Infra-red camera |
| Pattern Stimulator | |
| Luminance | 0-200 cd/m2 |
| Contrast | > 98% |
| Pattern Shape |
Checkerboard, Horizontal/vertical stripes, rectangle, adjustable element angle |
| Stimuli Mode | Reversal, On/Off, String |
| Stimuli Field | Full field, left/right/upper/lower half, up-down/left-right margin; |
| ¼ left-top/right-top/left-bottom/right-bottom; | |
| horizontal/vertical center, 1/8 center; | |
| Center circle with any angle; | |
| Field outside center circle with any angle. |

Advanced Clinical and Reserach Multifocal Visual Electrophysiology Device...

Integrated Clinical Visual-field VF Defect Visual Electrophysiology ...

Visual Field Perimetry Test Analyzer Automatic Ophthalmic Perimeter...

OLCR Optical Ophthalmic Biometer for Visual Axis & IOL Calculation...